Quick Start Guide
11
December 2014
Wiring 0-10 volts inputs on the voltage adapter
Wiring voltage 0-10 volt inputs using the adapter follows the same procedure as
mV inputs and thermocouples.
Figure below shows how to connect the voltage leads.
Adapter requirements
1. The adapter is only designed to be used with the 1000 mV sensor type, found
on device revisions 3 and above. If it is ordered pre-installed from the factory,
this will be the default sensor type. If the adapter is ordered as a spare part, the
user must configure the inputs to this sensor type. The user is responsible for
converting the 0-1000 mV transmitter output into a 0-10 volt scale. The
formula to do this is as follows:
2. If input type S004 ((1) dual channel voltage adapter) is ordered, it will be
factory installed on channels 1 and 2. However, if the adapter is required to be
installed on channels 3 and 4, the procedure to do so is a simple process.
Confirm that channels 3 and 4 are configured for 1000 mV sensor input. After
confirmation, remove the adapter from channels 1 and 2 and follow the steps
provided in the 'Installing the Optional Voltage Adapter' section of this guide
to install it on channels 3 and 4.
3. In order to ensure the device remains within the accuracy specifications, the
effect of source impedance must be checked.
Loaded to unloaded, the impedance ratio cannot exceed 0.1%. For detailed
instructions on how to verify this, refer to Section 5 of the User’s Manual.