Configuration
FD_ OFFSPEC _MASK
The FD_OFFSPEC_MASK parameter will mask any of the failed conditions listed in
FD_OFFSPEC_MAP. A bit on means the condition is masked out from alarming and will not be
reported.
FD_ OFFSPEC _PRI
FD_OFFSPEC_PRI designates the alarming priority of the FD_OFFSPEC _ALM, see Table 3-4 on
page 31. The default is 0 and the recommended values are 3 to 7.
FD_ OFFSPEC _ACTIVE
FD_ OFFSPEC _ACTIVE parameter displays the active alarms is active that are being selected for
this category. Only the alarm with the highest priority will be displayed. This priority is not the
same as the FD_ OFFSPEC _PRI parameter described above. This priority is hard coded within the
device and is not user configurable.
FD_ OFFSPEC _ALM
An alarm indicating the device needs maintenance soon. If the condition is ignored, the device
will eventually fail. FD_ OFFSPEC _ALM parameter is used primarily to broadcast a change in the
associated active conditions, which are not masked, for this alarm category to a Host System.
FD MAINT ALARMS
A maintenance alarm indicates informative conditions that do not have a direct impact on the
device's primary function(s).There are five parameters associated with MAINT_ALARMS, they
are described below.
FD_MAINT_MAP
The FD_MAINT_MAP parameter contains a list of conditions that do not have a direct impact on
the device's primary function(s).
Table 3-10. Maintenance Alarms/Priority Alarm
FD_MAINT_MASK
The FD_MAINT_MASK parameter will mask any of the failed conditions listed in
FD_MAINT_ENABLED. A bit on means that the condition is masked out from alarming and will
not be reported.
Sensor 3 Degraded 5
Sensor 4 Degraded 6
Sensor 5 Degraded 7
Sensor 6 Degraded 8
Sensor 7 Degraded 9
Sensor 8 Degraded 10
Alarm Priority
Excessive Deviation 1
Excessive Rate of Change 2
Table 3-9. FD Offspec Alarms