49
Reference Manual
00809-0100-4697, Rev GA
Section 3: Configuration
November 2014
Configuration
Table 3-21. Transducer Status Sub-parameter Structure
Table 3-22. DUAL_SENSOR CONFIG Sub-parameter Structure
Table 3-23. DUAL_SENSOR_STATUS Sub-parameter Structure
Transducer status table
0x01 A/D Failure
0x02 Sensor Failure
0x04 Dual Sensor Failure
0x08 CJC Degraded
0x10 CJC Failure
0x20 Body Temp Failure
0x40 Sensor Degraded
0x80 Body Temperature Degraded
Parameter Description
DUAL_SENSOR_MODE Disables or enables a sensor for configuration
DUAL_SENSOR_TAG Differential description
INPUT_A Sensor to be used in DUAL_SENSOR_CALC
INPUT_B Sensor to be used in DUAL_SENSOR_CALC
DUAL_SENSOR_CALC
Equation used for the dual sensor measurement including: Not Used,
Difference (Input A - Input B), and Absolute Difference (Input A - Input B)
ENG_UNITS Units used to display sensor parameter
UPPER_RANGE Upper Differential Limit (Input A High - Input B Low)
LOWER_RANGE Lower Differential Limit (Input A Low - Input B High)
Dual sensor status table
0x00 Active
0x01 Out of Service
0x02 Inactive
0x04 Component Sensor Open
0x08 Component Sensor Short
0x10 Component Sensor Out of Range or Degraded
0x20 Component Sensor Out of Limits
0x40 Component Sensor Inactive
0x80 Configuration Error